Maintenance and Calibration
81331 Rev 01 Nov 2010
99
5. With the main breaker OFF, plug the input power cord into the test set
AC POWER receptacle and into a three-wire grounded power receptacle
having the appropriate voltage and current ratings.
6. Suspend the outboard end of the high-voltage cable in free air so that it is
clear of all surrounding objects by at least 3 ft (0.91 m). Use dry nylon
rope if available.
7. Start your Delta4000 and use Delta Control (or a PowerDB form) and
perform a general GST-GND test.
When test is completed, observe test results. The results should be as follows:
Capacitance: between 4.0 and 8.0 pF
%DF or %PF: between -1.0 to +2.0%
Watts @ 10 kV: between -0.002 to +0.006
mA @ 10 kV: between 0.015 to 0.030
Repair
Megger offers a complete repair service and recommends that its customers take
advantage of this service in the event of equipment malfunction. Please indicate
all pertinent information including problem, symptoms, and attempted repairs.
Pack the Delta4000 in its transit case and include all cables that came with the
instrument. Equipment returned for repair must be shipped prepaid and insured
and marked for the attention of the Repair Department.